Procedure 3.1 Accumate RDO (Sigma Aldrich, www.sigmaaldrich.com, 800-325-3010) Rapid Bone Decalcification Procedure
Recipes for Microwave Staining Kits
SUMMARY
This procedure is intended for use as a "stat" alternative
to room temperature decalcification of compact bone specimens (e.g.
femur head). Due to variability in bone density from one specimen
to the next, 2-3 repeats of the final irradiation step may be necessary.
It is critical to perform flexibility checks at the stated intervals
to prevent over-decalcification and loss of nuclear detail. For optimal
effectiveness, the specimens should be secured in a microwavable
cassette rack and placed in 300 ml of Accumate RDO Rapid Decalcification
Solution in a 600 ml microwavable container. Cassettes should be
positioned in the rack so that they are located in the mid-portion
of the Accumate RDO Rapid Decalcification Solution during the procedure.
Agitation of cassettes between irradiation steps is a critical factor
in penetration and performance of the decal solution. Bone specimens
used should be no greater than 5 mm in thickness.
Note: This procedure is not intended for use on Cortical Bone (e.g.
bone marrows), since cortical bone may be rapidly decalcified in
Accumate RDO Decalcification Solution in 30-45 minutes.
Note: Bone must be fixed prior to decalcification with 10% Neutral
Buffered Formalin.
PROCEDURE
- Trim fixed bone specimen to 2-5 mm in thicknes, and place in
a plastic cassette. Use one cassette per specimen and secure cassette(s)
in a microwavable rack.
- Pre-heat 300 ml of Accumate RDO Rapid Decalcification Solution
at 800 watts for 25 seconds, without the rack/cassettes containing
the tissue.
- Place rack/cassettes containing tissue into the preheated Accumate
RDO Rapid Decalcification Solution and agitate rack up and down
20 times. (Use air bubble agitation if available).
- Place container containing rack/cassettes in the microwave cavity
and microwave at 200 watts for 15 minutes.
- Take container out of cavity in a well-ventilated area or under
a fume hood and agitate rack/cassettes up and down 20 times.
- Place container containing rack/cassettes in microwave cavity
and microwave at 300 watts for 15 minutes.
- Take container out of cavity in a well-ventilated area or under
a fume hood and agitate rack/cassettes up and down 20 times.
- Place container containing rack/cassettes in microwave cavity
and microwave at 400 watts for 10 minutes.
- Take container out of cavity in a well-ventilated area or under
a fume hood and agitate rack/cassettes up and down 20 times.
- Place container containing rack/cassettes in microwave cavity
and microwave at 200 watts for 10 minutes.
- Take container out of cavity in a well-ventilated area or under
a fume hood and agitate rack/cassettes up and down 20 times. Determine
decalcification endpoint by checking bone using standard methods.
If decalcification is not complete, repeat final irradiation step
(#10) until desired decalcification endpoint is acheived.
Note: It is critical to check the bone specimen
prior to each repetition of final irradiation to prevent over-decalcification
and loss of nuclear detail.
Note: This procedure should only be used in a laboratory
microwave with a ventilation system. |